Step 1: Air Supply Connection
- Connect an air source (filtered, dry air) to the 8 mm push-in air inlet.
air filter
To prevent particles to enter the air line, and potentially damaging the valves/flowmeters a filter with a maximum particulate size of 50 microns is recommended.
- Adjust the pressure regulator to 2 bars.
maximum pressure
The absolute maximum air input pressure is 115 psig / 8 bar.
Step 2: Clock Synchronization
- If using a Harp clock synchronizer board, connect the correspondent audio cable to the clock sync (CLK SYNC) input.
Step 3: Power Connection
- Plug the 24V DC power adapter into the barrel connector at the back panel, ensuring that the air supply is active before powering on the device.
- The green LED in the front panel must start blinking.
Step 4: USB connection
- Connect the USB cable to the back panel USB mini port and to your computer.
Step 5: External Control Setup
- If using external TTL control, connect the pluggable screw terminal to the back panel header.